1.- Important

Read the manual carefully before using this product and keep it in case of need.

If you cannot read or understand any warning, caution or instruction, contact the seller before using the product or you may be at risk of harm.

Consult your physician or orthopaedic specialist before use.

2.- Icons used on the label and in the manual
datos del fabricante Manufacturer details fecha de fabricacion Date of manufacture
dispositivo medico (medical device) MD Medical Device precaución Observe safety instructions
peso maximo kg Maximum user weight in kg CE Declaration of Conformity
atrapamiento Caution, danger of entrapment manual de usuario Check the user manual
reciclaje Product recycling information punto limpio Recycling centre
SN Product serial number LOT Product batch number
REF Product reference or code
datos del fabricante Manufacturer details
fecha de fabricacion Date of manufacture
dispositivo medico (medical device) MD Medical Device
precaución Observe safety instructions
peso maximo kg Maximum user weight in kg
CE Declaration of Conformity
atrapamiento Caution, danger of entrapment
manual de usuario Check the user manual
reciclaje Product recycling information
punto limpio Recycling centre
SN Product serial number
LOT Product batch number
REF Product reference or code
3.- Product use

Wheelchairs are aids that allow people to move about freely while seated without having to use the lower limbs.

4.- Product description
despiece silla

Model identification

Model Description
Lider R600 Lightweight manual wheelchair made of steel with fixed backrest. 600 mm rear wheels
Lider R300 Lightweight manual wheelchair made of steel with fixed backrest. 300 mm rear wheels
5.- Warnings and safety
precaución

Failure to follow these safety precautions may result in serious injury.

6.- Assembly and use

Assembly and folding

The wheelchair does not need to be assembled before use except for the footrests, which are shipped unattached to the wheelchair. No tools are required for the assembly of the footrests and armrests.

To place the footrests, follow these steps:

  1. Position the footrest in its place on the chassis, with the plate facing the outside of the chair, and then insert both footrest plates into their corresponding axes on the chair.
  2. Rotate the footrest, bringing the plate to its normal use position. At that moment, the lever will click, and the footrest will be fixed.

To adjust the height, unscrew the plate from the tube, place it at the desired height in one of the 5 different settings, and screw it back in to fix the height.

To remove the footrests, follow these steps:

  1. Pull the footrest lever, pushing it outwards from the chair, until the plate is on the outside of it.
  2. Pull up on the footrest to remove it.

Placement and removal of the armrests:

To fold and unfold the chair, follow these instructions:

User adaptation

Users of lightweight manual chairs should consider three very important things when choosing a chair.

First, the height and weight of the intended user, i.e. a wheelchair must always fit its user. Second, be clear what you are going to do in the wheelchair (transfers, folding, etc.). Finally, what is the environment in which the wheelchair is going to be used like.

The table makes recommendations that should be taken into account when choosing a wheelchair to meet the needs of a particular user:

USER MEASUREMENT RELATED TO WHEELCHAIR MEASUREMENT
Arm length Push rim reach Angle between arm and forearm 120° when holding the push rim
Head-to-floor height Sight distance
Height elbows-floor Armrests height 20mm above elbow with arms extended
Knee height Footrest-to-seat distance Footrest height 100-130mm
Distance buttock-back of knee Distance footrest-seat Seat Clearance of 25mm between thighs and side of armrest
Leg length Seat depth 30-50mm between the seat and the back of the knee
Hip width Seat width Clearance of 25mm between thighs and side of arm rest
Knee angle Seat and footrest tilt Seat tilt 1-4° backwards
Distance armpit-outside of hip Backrest height Backrest 25 mm below the shoulder blade
Approximate weight Wheelchair resistance Maximum user weight of 140 kg

Using the wheelchair

This section provides basic recommendations for use of the wheelchair by the person using it and any companion:

Use of the attendant brakes option

The attendant brakes allow the wheelchair to be stopped safely and conveniently by a companion.
Warning: Brakes should only be installed by operatives authorised by Forta.
Press the trigger on each of the wheelchair’s handles to operate the brakes. To lock the brakes and leave them on, push the red component on the front of the lever.

Freno acompañante
7.- Technical Specifications

The measurements and weights stated in this manual are for the standard wheelchair and may vary for a wheelchair with accessories or options. Changing wheels or making modifications may mean that the wheelchair does not comply with its type approval.

The specific measurements for each of the model in the wheelchair range are set out below. All measurements are in millimetres (mm) except for weight which is in kilograms (kg). Units not shown in the table for ease of reading and lack of space.

A
Height
B
Length
C
Seat height
D
Seat depth
E
Rear wheel diameter
F
Front wheel diameter
G
Length without footrests
H
Width
I
Backrest fabric height
J
Seat width
K
Weight
Líder R600
Talla 45 48
A 920 920
B 1100 1100
C 510 510
D 420,5 420,5
E 600 600
F 200 200
G 810 810
H 630 660
I 440 440
J 430 460
K 15,00 15,10

Líder R300
Talla 45 48
A 920 920
B 1070 1070
C 510 510
D 420,5 420,5
E 300 300
F 200 200
G 780 780
H 580 610
I 440 440
J 430 460
K 13,10 13,20

8.- Care and maintenance
precaución

If the product is not cared for and maintained in perfect condition, there will be a risk of accident.

Regular preventive maintenance allows you to keep your wheelchair in optimal working condition and to detect and solve small problems that could develop into major failures over time. Not only does it improve day-to-day safety, but it also saves time and money in the long run.
Note: In the event of a breakdown, any repair should be carried out by a specialist provider (orthopaedics).

General recommendations
Some of the most frequent maintenance and care for your wheelchair are described below:

ACTION FREQUENCY
Unfolding of the chair and its accessories Daily
Checking of nuts and bolts Weekly
Inflatable wheels Weekly
Condition of push rims Weekly
Brake condition Daily
Cleaning Daily
Presence of cracks Weekly
Lubrication Weekly
Upholstery condition Weekly
Disassembly mechanism Monthly
Adjustable components Monthly
Bearings Monthly
9.- Recycling the product
reciclaje punto limpio

When the product has become unusable and requires disposal, contact your specialist dealer. If you want to recycle it yourself, ask your local waste management company or nearest recycling centre for advice on disposal.

10.- Storing the product
precaución

If not correctly stored, the product may be damaged leading to the risk of an accident.

  • Do not store this product for an extended period of time near a heat source or in the sun (e.g. in front of a window or near a radiator) or near a cold source.
  • Keep the product away from flames and sparks.
  • Store the product as follows:
    • Store the product in a dry, cool place.
    • Protect the product from dust and rust by packaging or bagging it.
    • Store all removed parts or components together in one place so they do not get mixed up with anything else or get lost.
    • Do not place any weight on any components (do not place heavy objects on the product or any component).
11.- Useful life

The expected useful life of the product is 1 year in normal conditions of use, safety and maintenance. After that time, the product can be used as long while it remains in perfect conditions of use and maintenance. Some parts or components may in any event have to be replaced due to normal wear and tear within less than this time.

12.- Warranty and after-sales service

Warranty terms and conditions:
Only original Forta products purchased from authorised dealers are covered by the warranty against defects in parts and workmanship.

The warranty runs from the date of purchase for 3 years for structure and components subject to the following conditions:

  • Only original FORTA products purchased from authorised dealers will be covered by the warranty against defects in parts and workmanship.
  • The purchasing establishment must give notice of any defect in the product immediately.
  • The manufacturer will not have any liability for defects caused by improper handling, negligence or failure to follow the instructions set out in this manual.
  • During the warranty period, any part that proves defective due to faulty assembly or materials will be repaired or replaced free of charge by Forta.
  • The warranty will be absolutely void if incorrect modifications or adjustments are made to the product or if parts or options not supplied for the product model are fitted.
  • Damage from rust caused by misuse or negligence or the action or any liquid or gas that may damage the product are excluded from the warranty.

THE WARRANTY DOES NOT COVER THE CONSEQUENCES OF ANY TRAVEL PROBLEM, IN PARTICULAR ISSUES RELATING TO TRANSPORT, LOSS OF PROFIT, EXPENSES, ETC.

Forta will have NO liability in the following circumstances:

  • Normal wear and tear of the product or inappropriate or incorrect use.
  • Defects in any assembly or repair carried out by the purchaser or a third party.
  • Negligent handling or maintenance or use of unsuitable parts.
  • Repairs or modifications carried out in non-official workshops.
  • Forta will have no liability for any harm or loss suffered by the user or any third party as a result of any modification to or alteration of any product.
  • The warranty also excludes modifications made other than with original Forta parts.

EThe warranty certificate must be sent to FORTA, S.L. together with a photocopy of the purchase invoice within 15 days from the date of purchase. To validate the warranty, the user must submit the warranty certificate bearing the stamp the seal of the outlet that sold the product. You can download the warranty certificate here: www.fortasl.es/en/manuales/doc/GARANTIA CONTRACTUAL.pdf.

13.- Tests and trials carried out

FORTA products comply with Annex IV to MDR (EU) 2017/745, are CE marked and have been designed and tested in accordance with UNE-EN ISO 12183:2014. Forta is obliged to certify compliance by its products with a number of essential requirements. There is a standard that specifies the inspection and testing methods to show compliance of the products with those requirements. The standards that the product complies with are listed below:

Brief description of relevant law Current law
Information provided by the manufacturer UNE-EN 1041:2009+A1:2014
Quality management system UNE-EN ISO 13485:2018
Risk management UNE-EN ISO 14971:2020
Testing of medical devices and manual wheelchairs UNE-EN 12182:2012
UNE-EN 12183:2019
Biological assessment of medical devices UNE-EN ISO 10993-1:2021
Fire resistance testing UNE-EN 1021-1:2015
UNE-EN 1021-2:2015
Maximum load
14.- Manufacturer details

FORTA - FABRICACIONES ORTOPÉDICAS ALBACETE, S.L.
Pq. Emp. Campollano, c/C 90 Nave 1
02007 Albacete
España - Spain
info@fortasl.es
www.fortasl.es
CE
MADE IN EUROPE
forta